Description

Standard reference force transducers specially designed according to the ISO 376 standard (classes 1, 0,5 and 00). o Wide range of capacities o Tension and / or compression (universal) o Compact design o Complete range of load accessories o Protection class: IP65 o Material: alloy steel (stainless steel available on request) o Cable length: see drawing (other lengths available on request) Technical drawings available on datasheet

Product features

Relative reversibility error (% M.V.**) <± 0.3
Relative repeatability error with rotation (% M.V.**) <± 0.2
Relative repeatability error without rotation (% M.V.**) <± 0.1
Stabilization time after power excitation supply (s) 200...600
Creep error over 30 min. (% F.S.*) <± 0.1
Reference temperature (°C) 20
Compensated temperature range (°C) -10...+45
Service temperature range (°C) -25...+70
Storage temperature range (°C) -50...+85
Temperature coefficient of the sensitivity (% F.S.*/10°C) <± 0.035
Temperature coefficient of zero signal (% F.S.*/10°C) <± 0.03
Zero balance (mV/V) ± 0.02
Nominal sensitivity (mV/V) 2
Input resistance (ohm(s)) 350 ± 2
Output resistance (ohm(s)) 350 ± 2
Insulation resistance (50 V) (Mohm(s)) > 5000
Reference excitation voltage (VDC) 10
Permissible nominal range of excitation voltage (VDC) 3...12
Safe load limit (% F.S.*) 110
Breaking load (% F.S.*) >300

Additional product literature