Precision PT100 simulator - Model 4506, 4506 S

Pt100 calibration simulator, for high precise checking and calibrating, compact
Web icon

Description

The precision Pt100 simulator is used wherever measuring instruments or controlling systems have to be tested or calibrated with great precision. The standard version model 4506 simulates 24 set temperature values. Totally different temperature ranges are required in the case of special applications, such as in the food industry, the medical science etc. Model 4506 S is recommended for these applications and can be fitted with 24 temperature values according to customer specifications. A distinctive feature of these simulators is their simple operation. The resistance values required for simulation are directly set in °C. There is no need to use loof-up tables. Today in many cases a certificate is needed more than ever force measurements. This certificate is available under order code 45DKD-.. . Features: 24 set temperature values Calibration per DIN EN 60751 Resistor material: MANGANIN®, TK ≤ 10 ppm/K Pt100, Pt500, Pt1000 Available with DKD Certificate if required

  • Calibration equipment
  • simulator
  • calibration
  • high precise

Product characteristics

Pt100 simulation range Model 4506
-100 °C bis +500 °C
Pt100 simulation range Model 4506S
Any value in the range between -100 °C ... +850 °C is possible.
Measurement accuracy
± (0.082 + 0.0003 x t) in °C
Pt200, Pt500, Pt1000 simulation
4506 - No / 4506 S - Pt500, Pt1000
Ni100, Ni1000 simulation
No
Supply voltage
Passive
Interfaces
No

Documents

Domain icon Manufacturer/ Producer

76593 Gernsbach - Germany

Contact